Easy Method Transfer from MicroCal VP-ITC to PEAQ-ITC

If you previously used a Microcal VP-ITC for binding studies, you can perform the same experiments on a MicroCal PEAQ-ITC. Use the experimental design wizard with the PEAQ-ITC software to predict ITC results with known parameters, and look at different concentrations and injection conditions.

Please login or register for free to read more.

Introduction

If you previously used a Microcal VP-ITC for binding studies, you can perform the same experiments on a MicroCal PEAQ-ITC. Use the experimental design wizard with the PEAQ-ITC software to predict ITC results with known parameters, and look at different concentrations and injection conditions.

Comparison of recommended ITC settings

SettingVP-ITCPEAQ-ITC
Temperature setting2-80 °C. Most experiments performed between 10-37 °C2-80 °C. Most experiments performed between 10-37 °C
Reference power10-30 µcal/sec3-10 µcal/sec
Stirring speedApprox. 300 rpmRecommended 750 rpm
Cell volume / volume needed to fill cell1.4 ml/2 ml200 µL/300 µL
Injection Syringe volume / volume needed to fill syringe300 µL/500 µL40 µL/75 µL
Range of injection volume5-15 µL1-3 µL
Typical injection conditions1st injection 2 µL
Injections 2-58: 5 µL
OR injections 2-29: 10 µL
OR injections 2-19: 15 µL
1st injection 0.4 µL
Injections 2-39: 1 µL
OR injections 2-19: 2 µL
OR injections 2-13: 3 µL
FeedbackHighHigh
Filter period2 secNA

Table 1: Comparison of recommended ITC settings

[Figure 2 TN240816-method-development-vp-itc-peaq-itc.jpg] Figure 2 TN240816-method-development-vp-itc-peaq-itc.jpg

Figure 1: PEAQ-ITC experiment simulator

Calculating sample concentration

Sample concentration depends on the binding affinity and the C parameter. 

Assuming N = 1: C value = [Concentration in ITC cell]/KD

  • Best between 5 and 500
  • If needed, the C value can be down to 1 or up to 1000

For PEAQ-ITC, the minimum recommended concentration is 5-10 µM. For large heat change and/or tight binding, you can reduce the concentration to 1-5 µM. Weak binding and/or low heat changes need higher sample concentrations.

Sample concentration in ITC syringe

For 1:1 binding, use 10X the concentration in the ITC cell.

Since the PEAQ-ITC cell is smaller, a binding experiment in the PEAQ-ITC produces less heat change/injection and total heat change, compared to same titration performed with a VP-ITC. In many cases, the fit binding curve produces similar KD and ΔH results with both instruments (see figure 2 and table 2 below).

NOTE: It has been observed that if the ΔH is low, you may need higher sample concentration in the PEAQ-ITC, compared to the VP-ITC to achieve high-quality results.

[Figure 1 TN240816-method-development-vp-itc-peaq-itc.jpg] Figure 1 TN240816-method-development-vp-itc-peaq-itc.jpg

Figure 2: Left: PEAQ-ITC experiment with 0.1 mM EDTA in cell, 1 mM CaCl2 in syringe. Injection 1 0.4 µL. Injections 2-20 2 µL. N=1.01, KD 573 nM, ΔH -3.91 kcal/mole. Heat change for injection 2: -7.9 µcal.  Total heat change: -78.2 µcal. Right: VP-ITC experiment with 0.1 mM EDTA in cell, 1 mM CaCl2 in syringe. Injection 1 2 µL. Injections 2-29 10 µL. N=1.03, KD 563 nM, ΔH -4.14 kcal/mole. Heat change for injection 2:  -40.5 µcal.  Total heat change: -562 µcal

Heat change guidelines for high-quality ITC data

Heat changeVP-ITCPEAQ-ITC
Integrated Injection heat for 2nd injection (first full injection)More than 10 µcal is ideal (minimum 5 µcal)More than 2.5 µcal is ideal (minimum 1 µcal)
Total heat change for all injectionsMore than 80 µcal – below this value data can be noisier and result in less reliable fitsMore than 16 µcal – below this value data can be noisier and result in less reliable fits

Table 2: Heat change guidelines for high-quality ITC data

ITC experiment design resources

For more guidelines on designing ITC experiments, please take a look at the resources below.

  1. Practical tips for MicroCal PEAQ-ITC experiments
  2. Best Practices for Isothermal Titration Calorimetry to study binding interactions – Part 1
  3. Best Practices for Isothermal Titration Calorimetry to study binding interactions – Part 2
  4. Best Practices for Isothermal Titration Calorimetry to study binding interactions – Part 3
  5. Best Practices for Isothermal Titration Calorimetry to study binding interactions – Part 4

ログイン

まだ登録されていませんか? アカウントを作成する